Evaluation of the diagnostic accuracy of Xpert® Mpox and STANDARD™ M10 MPX/OPX for the detection of monkeypox virus
Romero-Ramirez A., Somasundaran A., Kontogianni K., Parkes J., Hussain Y., Gould S., Williams CT., Wooding D., Body R., Hardwick HE., Baillie JK., Dunning J., Semple MG., Fletcher TE., Edwards T., Emperador D., Cubas-Atienzar AI.
Objectives: Evaluation of diagnostic accuracy of two point-of-care (POC) molecular diagnostic tests for the detection of monkeypox virus (MPXV): Xpert® Mpox (Cepheid, Inc., USA) and STANDARD™ M10 MPX/OPX (SD Biosensor, Inc., Korea). Methods: Diagnostic accuracy of both POC platforms was evaluated using 53 upper-respiratory swabs (URS) and 32 skin lesions swabs (SS) collected from mpox and COVID-19 patients in the UK against the Sansure (Sansure Biotech Inc.) and CDC reference qPCR tests. The analytical sensitivity of both platforms was assessed using a viral isolate from II, B.1 lineage. Results: The overall sensitivity and specificity of the Xpert® Mpox was 97.67% [95% CI 87.71–99.94%] and 88.57% [95% CI 73.26–96.80%] and 97.44% [95% CI 86.52–99.94%] and 74.42% [95% CI 58.83–86.48%] comparing the Sansure and CDC qPCR, respectively and for the M10 MPX/OPX was 87.80% [95% CI 73.80–95.92%] and 76.60% [95% CI 61.97–87.70%] and 94.29% [95% CI 80.84–99.30%] and 86.67% [95% CI 73.21–94.95%] with the Sansure and CDC qPCR. Conclusion: The Xpert® Mpox had good diagnostic accuracy for both sample types while the M10 MPX/OPX clinical accuracy was deficient with URS. Our data supports the use of URS during the first 3 days of symptoms onset for mpox diagnosis.